Chapter 2: Control Testing

Why Run Control Tests

Control tests ensure that both the meter and the test strips are working properly and giving reliable results.
You should run a control test when:
The meter is new
You open a new box of test strips
You leave the cap off of the test strip container
You want to check the performance of the meter
and test strips

About the Control Solution

Make sure you use only ACCU-CHEK Active
control solution.
When you first open the bottle, write the date on
the label – the control solution is good for three
months from that date or until the expiration
date on the bottle, whichever comes first.
You drop the meter
Your test results don't agree with how you feel
You want to make sure you're performing tests
correctly
Never use control solution that has expired.
Always close the control solution bottle tightly.
Store at room temperature, between 36-86°F
(2-30°C).